    Wood MH, Sampson DC, Hensley WJ.
    Clin Chim Acta. 1977 Jun 15;77(3):343-7.
    A gas chromatographic method for the plasma assay of the anticonvulsant, sodium valproate, is described. Derivatization is not necessary. 200 microliter plasma are required for a single estimation. The method involves a chloroform extraction of valproate and the internal standard, cyclopentane carboxylic acid, from acidified plasma. Gas-liquid chromatography using the stationary phase 10% SP-216-PS gives complete separation of valproate and the internal standard in eight minutes. The limit of detection is 20 mumol valproate/1 plasma (equivalent to 40 pmol on column). This is well below the lower therapeutic plasma level. The between-run precision of the method indicates a variation for each sample within+/-3% of its mean value.